Bard 2-Way Foley Catheter Features

  • This uncoated foley catheter reduces friction and improves patient comfort
  • Latex-free
  • Sterile
  • For single patient use

What to buy with Bard Two-Way Uncoated Foley Catheter

  • Bard Bardia Closed System Urinary Drainage Bag
  • Bard StatLock Foley Stabilization Device – Adult
  • Medline Urinary Drain Bag Cover

Bard Two Way Indwelling Catheter Inflation/Deflation

  • Catheter Inflation:
    • Ensure that the Bard Foley catheter balloon is positioned well within the patients bladder. Slowly, with a gentle, constant force, inflate the Bard Foley catheter balloon with the volume prescribed on the package
    • Note that the 5cc balloon must be inflated with between 9cc to 10cc of sterile water
    • Improperly inflated Bard Foley catheter balloons may cause drainage and deflation difficulties
    • Using fluid other than sterile water may cause the balloon not to empty properly, especially after long dwell times
  • Catheter Deflation:
    • Select a luer slip syringe
    • Slide the plunger of the syringe up and down the barrel of the syringe several times to loosen it up
    • Compress the plunger all the way and then pull back the plunger slightly so that it does not adhere to the front of the syringe barrel
    • Gently insert the syringe in the catheter valve
    • Do not use more force than is required to make the syringe stick in the valve
    • Allow the pressure within the balloon to force the plunger back and fill the syringe with water
    • If you notice slow or no deflation, re-seat the syringe gently. Once again, allow the balloon to deflate slowly on its own
    • If the balloon does not deflate, reposition the patient.
    • Ensure that the catheter is not in traction, the proximal end of the catheter is not compressed within the bladder neck
    • Ensure that urine flows freely
    • Attempt to deflate the balloon by using the pressure in the balloon to force water into the syringe as described above
    • If the balloon stills fails to deflate, apply very gentle slow aspiration. Aspiration that is too rapid, or too forceful, may cause the inflation lumen within the Foley catheter to collapse
    • If permitted by hospital protocol, the valve arm may be severed
    • If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol